ANALYTICAL LETTERS, cilt.42, sa.18, ss.2978-2985, 2009 (SCI-Expanded)
This paper describes two rapid, sensitive, and specific methods for the determination of medazepam with internal standard (diazepam) in pharmaceutical preparations by gas chromatography (GC) with flame ionization (FID) and mass spectrometric (MS) detection. The linearity was established over the concentration range of 0.25-5 mu g/mL for GC-FID and 50-1000ng/mL for GC-MS method. The intra- and inter-day relative standard deviation was less than 2.94 and 4.95%, respectively. Developed GC-FID and GC-MS methods in this study are accurate, sensitive, and precise and can be directly and easily applied to Tranko buskas as pharmaceutical preparation.