JOURNAL OF CHROMATOGRAPHIC SCIENCE, cilt.53, sa.9, ss.1443-1448, 2015 (SCI-Expanded)
A simple high-performance liquid chromatography method has been developed for determination of flurbiprofen in human plasma. The method was validated on an Ace C-18 column using UV detection. The mobile phase was acetonitrile-0.05 M potassium dihydrogen phosphate solution (60: 40, v/v) adjusted to pH 3.5 with phosphoric acid. The calibration curve was linear between the concentration range of 0.10-5.0 mu g/mL. Intra- and inter-day precision values for flurbiprofen in plasma were < 4.47, and accuracy (relative error) was better than 3.67%. The extraction recoveries of flurbiprofen from human plasma were between 93.0 and 98.9%. The limits of detection and quantification of flurbiprofen were 0.03 and 0.10 mu g/mL, respectively. In addition, this assay was applied to determine the pharmacokinetic parameters of flurbiprofen in six healthy Turkish volunteers who had been given 100 mg flurbiprofen.