Non-enzymatic electrochemical glucose sensor based on CeO2–MoO3 decorated laser-induced graphene


ORHAN Z., ACAR M., ÖZTÜRK DOĞAN H.

Chemical Physics Letters, vol.897, 2026 (SCI-Expanded, Scopus)

  • Publication Type: Article / Article
  • Volume: 897
  • Publication Date: 2026
  • Doi Number: 10.1016/j.cplett.2026.143000
  • Journal Name: Chemical Physics Letters
  • Journal Indexes: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Chemical Abstracts Core, Chimica, Compendex, INSPEC, Academic Search Ultimate (EBSCO), Engineering Source (EBSCO)
  • Keywords: Electrochemical sensor, Laser-induced porous graphene, Flexible sensor, Glucose, CeO2, MoO3
  • Ataturk University Affiliated: Yes

Abstract

Herein, we demonstrate the fabrication of porous laser-induced graphene (LIG) electrodes decorated with both CeO2 and MoO3 nanostructures via sequential electrochemical deposition. It exhibited a linear glucose response between 0.1 and 1.8 mM, with a sensitivity of 2957.5 μA mM−1 and LOD of 1.52 μM; the sensor showed negligible interference from dopamine, uric acid, ascorbic acid, and various inorganic ions, whereas reproducibility was found to be in the range from 91.5% to 102.8%. These results highlight the potential of the CeO2–MoO3/LIG platform as a flexible and efficient electrode architecture for enzyme-free glucose sensing applications.