Improving the limit of detection (LOD) of microsensor used in detection of brain diseases via wavelet filter Beyin Hastaliklarinin Tanisinda Kullanilan Mikrosensörün Dalgacik Filtre Kullanilarak Algilama Alt Sinirinin Iyileştirilmesi


Koc H., Kadihasanoglu G., Geyikoglu M. D., DERTLİ M. E., ÇAVUŞOĞLU B., ÖZBEK İ. Y., ...Daha Fazla

2015 23rd Signal Processing and Communications Applications Conference, SIU 2015, Malatya, Türkiye, 16 - 19 Mayıs 2015, ss.1676-1679 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Cilt numarası:
  • Doi Numarası: 10.1109/siu.2015.7130174
  • Basıldığı Şehir: Malatya
  • Basıldığı Ülke: Türkiye
  • Sayfa Sayıları: ss.1676-1679
  • Atatürk Üniversitesi Adresli: Evet

Özet

© 2015 IEEE.Limit of detection (LOD) gives the concentration amount that a microsensor can detect. It is desirable to have a LOD value of 1μM for microsensors used in brain diseases. The ones that cannot reach this sensitivity value are disposed and cannot be used in the experiments. The goal of this study is to increase the sensitivity of the produced microsensors by decreasing their LOD values. LOD increases linearly by baseline noise. The sensor data is used generally without any baseline filtering in the literature. In this study, LOD values are enhanced 3 times as much by using wavelet filtering, compared with the ones where no filtering is used.