CHEMICAL INDUSTRY & CHEMICAL ENGINEERING QUARTERLY, cilt.19, sa.2, ss.313-319, 2013 (SCI-Expanded)
A novel analytical method was developed and validated for determination of prilocaine HCl in bulk drug and pharmaceutical formulation by gas chromatography-nitrogen phosphorus detection (GC-NPD). The chromatographic separation was performed using a HP-5MS column. The calibration curve was linear over the concentration range of 40-1000 ng ml(-1) with a correlation coefficient of 0.9998. The limits of detection (LOD) and quantification (LOQ) of the method were 10 and 35 ng m(-1), respectively. The within-day and between-day precision, expressed as the percent relative standard deviation (RSD%) were less than 5.0%, and the accuracy (percent relative error) was better than 4.0%. The developed method can be directly and easily applied for determination of prilocaine HCl in bulk drug and pharmaceutical formulation using internal standard methodology.